Effects of stimulus level and rate on psychophysical thresholds for interleaved pulse trains in cochlear implants

cm_sbs_024_plain.png

This study examined channel interactions using interleaved pulse trains to assess masking and potential facilitative effects in cochlear-implant recipients using clinically relevant stimuli. Psychophysical thresholds were measured for two adjacent mid-array electrodes; one served as the masker and the other as the probe. Two rates representative of those found in present-day strategies were tested: 1700 and 3400 pulses per second per channel. Four masker levels ranging from sub-threshold to loud-but-comfortable were tested. It was hypothesized that low-level maskers would produce facilitative effects, shifting to masking effects at high levels, and that faster rates would yield smaller masking effects due to greater stochastic neural firing patterns. Twenty-nine ears with Cochlear or Advanced Bionics devices were tested. High-level maskers produced more masking than low-level maskers, as expected. Facilitation was not observed for sub-threshold or threshold-level maskers in most cases. High masker levels yielded reduced probe thresholds for two Advanced Bionics subjects. This was partly eliminated with a longer temporal offset between each masker-probe pulse pair, as was used with Cochlear subjects. These findings support the use of temporal gaps between stimulation of subsequent electrodes to reduce channel interactions.