[Loudness optimized registration of compound action potential in cochlear implant recipients].
Laryngorhinootologie. 2017 Sep 22;:
Authors: Berger K, Hocke T, Hessel H
Abstract
Background Postoperative measurements of compound action potentials are not always possible due to the insufficient acceptance of the CI-recipients. This study investigated the impact of different parameters on the acceptance of the measurements. Methods Compound action potentials of 16 CI recipients were measured with different pulse-widths. Recipients performed a loudness rating at the potential thresholds with the different sequences. Results Compound action potentials obtained with higher pulse-widths were rated softer than those obtained with smaller pulse-widths. Conclusions Compound action potentials measured with higher pulse-widths generate a gap between loudest acceptable presentation level and potential threshold. This gap contributes to a higher acceptance of postoperative measurements.
PMID: 28938498 [PubMed - as supplied by publisher]
from #ENT-PubMed via ola Kala on Inoreader http://ift.tt/2fMzxbd
Δεν υπάρχουν σχόλια:
Δημοσίευση σχολίου